Innovative Technologies

POCT Immunofluorescence Quantitative Analyzer (QTX-POT329)

POCT Immunofluorescence Quantitative Analyzer (QTX-POT329)

Description

The POCT Immunofluorescence Quantitative Analyzer is used cooperatively with special fluoroimmunoassay reagent strips and applicable to fluoroimmunoassay of human body samples. The product can be used for semiquantitative and quantitative immunoassay of medical, antiepidemic and scientific research institutions.
  • Test Speed: 240 times/hour
  • WavelengthRange:740nm
  • Excitation LightSource:LED
Features
  • Test Speed: 240 times/hour
  • WavelengthRange:740nm
  • Excitation LightSource:LED
Technical Specification
Model QTX-POT329
Test Item The product is used cooperatively with immunofluorescence assay kit produced by our company.
Excitation Light Source LED
Wavelength Range 740nm
Test Item Parameter The user can read configuration parameters of the instrument from information card.
Sample Type Whole blood, serum, plasma, urine and faeces (subject to specification of immunofluorescence assay kit)
Delay Test Delay test time can be set. Place test card in the instrument and test at fixed time after completion of reaction
Test Speed 240 times/hour
Background Number Fluorescence value should be ≤5
Repeatability The variable coefficient of fluorescence value of fluorescence test cards of different intensities repeatedly measured by the instrument CV≤2.0%
Stability The relative deviation between the test result at 4th hour and the

8th hour after the instrument is working stably and the test result when the instrument just works stably is limited to ±3.0%.

Accuracy When the instrument tests the calibrated fluorescence test card, the deviation between the measured fluorescence value and the calibration value of the fluorescence test card is ±3.0%.
Measuring Time The time from inserting the fluorescence test card to displaying the test result should not exceed 15s.
Linearity Range Within the concentration range of standards no less than two orders of magnitudes, the linearly dependent coefficient of the fluorescence value and concentration read by the instrument r≥0.990
Electricity AC 100V-240V 50Hz/60Hz; Host: DC12V 5A
Input Power 60 VA
Dimension 200*300*140mm
Net Weight 2.0kg