Innovative Technologies

Kjeldahl Digestor (QXK Series)

Kjeldahl Digestor (QXK Series)

Kjeldahl Digestor adopts advanced high-temperature infrared tradiation heating technology and microprocessor control platform, boasts accurate temperature control and quick temperature rise. It has two kinds of temperature rise modes: linear and curve temperature rise mode, and offers 25 digestion programs for control of temperature rise curves. Neutralization system has many functions such as triple filtration, condensate recovery of exhaust gas, filtration and neutralization device. The product adopts high-quality anticorrosive pumps, low noise, strong suction, reduce exhaust emissions, eco-friendly.
Features

Gas Collection Hood

Using PFA sealed cap, longer life, dealing effect well.

Clip-on sealing cap, easy for changing.

Specialized water jet vacuum pump is adopted, don’ t need electric power.

Drip tray design, reduce corrosion damage from acid solutions.

 

Characteristics:

Can digest 20 samples at the same time

Using infrared radiation graphite conduction heating method, the heating speed is fast, and the temperature difference between the holes is small

Adopt PID temperature control technology, high temperature control accuracy, fast heating speed, only 25 minutes from room temperature to 400℃

It has two heating methods: curve heating and linear heating, 500 groups of digestion programs can be edited and stored, and each group can be set up to 25 segments

 

Temperature and time gradient to meet different experimental needs of users

The surface of the instrument is coated with Teflon, and the whole machine is treated with anti-corrosion treatment to avoid the corrosion of the machine by acidic liquids during the experiment

Using 5.6 inch large screen true color LCD display system, intuitive and convenient operation

The whole machine has multiple protections such as over-voltage, over-current, over-heating, etc., and over-temperature alarm function

Adopt advanced ceramic and air duct heat insulation technology to make the experiment operation safer

Standard simple cooling rack, digestion tube can be cooled quickly, improve work efficiency

Optional waste gas neutralization system, through condensation, neutralization and drying to eliminate the acid gas generated during the experiment

Technical Specification
QXK-D470

QXK-D480

Temperature range Room temperature +5 – 450℃

·

Temperature accuracy ±1℃

·

Heating method Infrared heating and high-purity graphite conduction

·

Heating insulation method Unique air duct insulation technology

·

Digestion tube capacity 300mL

·

Capacity per batch 20pcs/batch

·

Power supply 220 VAC±10% 50HZ

·

Power 3600W

·

Dimensions 515mmX458mmX730mm

515mmX421mmX211mm

Net weight 40Kg

25Kg