QXS-T167
Technical parameters
Heating Unit 8 pcs
Temperature Range RT ~ 85℃
Temp range of Waste tray RT ~ 50℃
Temperature of Cold Plant about -5℃
Temperature Control Precision ±1℃
Flow range of Paraffin nozzle 0 ~ 400 ml/minute
Lighting brightness Stepless control
Paraffin Chamber Capacity 6L 524 x 110 x 110 mm
Thermal Chamber Capacity 2.3L 247 x 182 x 55 mm
Dimensions of Work Plant 558 x 110 mm
Dimensions of Cold Plant 70 x 55 mm
Embedding hammer size 10×10, 7×12, 15×15, 20x20mm (Optional)
Temperature Ranges of Cryo Module 5 ~ -20℃
Dimensions of Cryo Module 395 x 390 mm
Carrying capacity of Cryo module ≦140 cassettes or 70 bottom mold
Electrical parameters
Working Voltage AC110 ~ 120V / AC 220 ~ 240V
Frequency 50Hz / 60Hz
Power 1300W
BM V Fuse specifications F10A L 250V / F20A L 125V
BL V Fuse specifications F6A L 250V / F12A L 120V
Electrical Safety I
Working environment
Ambient Temperature 5℃~45℃
Relative humidity ≤80%
Air pressure range 700hPa~1060hPa
Packaging parameters
Overall Dimensions 1020 x 675 x 373 mm
Net Weight 60kg
Quantix Equipment is a rapidly growing leader in innovation and quality, specializing in the development and supply of advanced pharmaceutical testing equipment and support services.